Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: A.W. Sperry DSA-2007: (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 67 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # 2007v_52.bmp # 20074z52.bmp # 2007lc52.bmp # 2007hc52.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 STD Fluke 5500A/COIL 1.005 HEAD PRELIMINARY INSTRUCTIONS 1.006 DISP Warm-up time: NONE 1.006 DISP Temperature: 25 +/- 5degC 1.006 DISP Humidity: <75% 1.007 DISP [32] WARNING 1.007 DISP HIGH VOLTAGE is used or exposed during the performance 1.007 DISP of this calibration. DEATH ON CONTACT may result if 1.007 DISP personnel fail to observe safety precautions. 1.008 HEAD {RESISTANCE} 1.009 DISP Set the FUNCTION/RANGE switch to OHMS. 1.010 PIC 20074z52 1.011 5520 190.0Z 1.5% 0.2U CW 2.001 5520 10Z S CW 2.002 EVAL Is an audible tone present {(Continuity test OK)}? 3.001 HEAD {AC VOLTAGE} 3.002 PIC 2007v_52 3.003 HEAD AC VOLTAGE: {750V Range} 3.004 5520 750 700V 1% 2U 60H SI 2W 4.001 HEAD AC VOLTAGE: {200V Range} 4.002 5520 200 100.0V 1% 2U 60H SI 2W 5.001 HEAD {AC CURRENT} 5.002 DISP Set the FUNCTION/RANGE switch to 200 AMPS. 5.003 PIC 2007lc52 5.004 HEAD AC CURRENT: {200A Range} 5.005 5520 200 100.0A 1.5% 0.4U 60H SI T5 6.001 HEAD AC CURRENT: {600A Range} 6.002 PIC 2007hc52 6.003 5520 600 500A 1.5% 2U 60H SI HC T5 7.001 END